Instrument Detail
 As a "National High-tech Enterprise" , Bioteke holds the philosophy to offer products with low price, high quality and superior services to the world, and insists on pursuing higher excellence.
Home » Products » Instrument Center » Nucleic Acid Extractor » Automated Nucleic Acid Testing Equipment Nucleic Acid DNA RNA Extraction Machine AU1001(Throughput-96)

Automated Nucleic Acid Testing Equipment Nucleic Acid DNA RNA Extraction Machine AU1001(Throughput-96)

For nucleic acid extraction from virus, tissue, cells, swab, faeces, plant, blood and other clinical or forensic samples
Availability:
Quantity:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
sharethis sharing button
  • AU1001-96

  • Bioteke

Description

BioTeke's Automated Nucleic Acid Extraction System uses cutting-edge magnetic bead separation technology to achieve fast and easy operation of DNA/RNA automation extraction procedure. 

All separation and purification procedures are completed in 96 well plates and there is no need for steps like regular centrifugation and filtering required by extraction and purification. 

The system can complete fast extraction and purification process from 1 to 32 samples, including but not limited to blood, tissue, cells, swab, virus and other clinical or forensic samples. 

The extracted high quality DNA/RNA is ready for use in highly sensitive downstream analysis, such as qPCR, clinical molecular diagnostic, analysis of gene expression, gene analysis, forensic and infectious disease research etc. 

The purified nucleic acid can be applied directly for the next enzyme cleavage, identification as well as disease diagnosis and treatment etc. As a famous domestic nucleic acid separation and purification technology company, BioTeke provides over 100 kinds of extraction kits for use with this instrument to maximize its application and minimize its costs.

 

This automated nucleic acid extractor uses magnetic rods on magnetic rod racks to transfer magnetic beads bound with nucleic acid to different wells, and mixes liquid at fast speed and repeatedly with magnetic sleeves outside magnetic rods. 

The liquid is mixed evenly. And after cell lysis, nucleic acid binding, washing and elution, highly purified DNA/RNA nucleic acid molecules is finally achieved.

 

 

Features


1. Large screen English operation panel:

Large screen, full color display, touch screen English operation panel, easy to use.

2. Precise control

Built-in engineering computer, no need to connect to PC. Single machine operation can help save more space and energy and provide highly stable automatic control system.

3. Temperature control

Customer can define lysis and elution temperature as needed.

4. Free programming

Powerful program editing function allows flexible and efficient defining of your application to meet requirements of various reagents.

5. High purity & high yield

The purification program can be optimized based on reagents. And with precise incubation time, higher extraction efficiency can be achieved. The highly purified DNA/RNA can be used directly for PCR and RT-PCR.

6. Stable results

Discrepancies and errors caused by manual operation can be avoided. The results are stable and come with good repeatability.


AU1001-96 Main Parameter


Model Type

AU1001-96

Product Type

Nucleic Acid Extraction

Product Name

Automatic Nucleic Acid Extractor

Dimensions

830mm*530mm*525mm

Weight

60KG

Power supply

220V

Consumables

Corresponding reagents,Model   No.AU17011,No. 52012

Operatir)g environment relative humidit

30%80%

Operating environment   temp

1530

Temperature control range

4 to ~100

Temperature control accuracy

±0. 1

Average heating rate

2.5 /s

Processing Volume

10-1000μL

Magnetic Force

4500gauss Above

throughput

96pcs at same time

Extraction   time

15min to 2hours, depend on Kit

Certification

CE, 13485, FDA

 

Testing data

Experiment 1 :FAM Channel Test Ct value

数据图1

 

Experiment 2:FAM Channel Test Ct Value

数据图2

 

According to above three testing data, 96 throughput RNA extraction system will not generate aerosol contamination, 

with obvious difference between medium & low concentration samples and positive & negative samples. High sample extraction uniformity, CV value less than 30%. 


Previous: 
Next: 
QUICK LINKS
PRODUCT CENTER
CONTACT US
Placeholder Image

   Telephone/Whatsapp: +86 18115363743

Placeholder Image

  Email : zr@bioteke.cn

Copyright BioTeke Corporation(wuxi) Co.,Ltd  |   苏ICP备18042459号-1